Can tape-screw fixation of a quadrupled semitendinosus graft in a full-length tibial tunnel provide superior fixation compared with a doubled semitendinosus-gracilis held with an interference screw? A matched-pair cadaveric biomechanical comparison

View/ Open
File version
Version of Record (VoR)
Author(s)
Vertullo, Christopher J
Cadman, Joseph
Dabirrahmani, Dane
Appleyard, Richard
Griffith University Author(s)
Year published
2018
Metadata
Show full item recordAbstract
Background:
In anterior cruciate ligament reconstruction, quadrupled semitendinosus (Quad ST) grafts have potential advantages over doubled semitendinosus–gracilis (ST/G) including larger diameter and gracilis preservation, however the ideal tibial fixation method of the resultant shorter Quad ST graft remains elusive if a fixed-loop suspensory fixation device is used on the femur. We investigated whether the tibial fixation biomechanical properties of a Quad ST fixed indirectly with polyethylene terephthalate tape tied over a screw in a full outside-in created tunnel was superior to a ST/G graft fixed with an interference ...
View more >Background: In anterior cruciate ligament reconstruction, quadrupled semitendinosus (Quad ST) grafts have potential advantages over doubled semitendinosus–gracilis (ST/G) including larger diameter and gracilis preservation, however the ideal tibial fixation method of the resultant shorter Quad ST graft remains elusive if a fixed-loop suspensory fixation device is used on the femur. We investigated whether the tibial fixation biomechanical properties of a Quad ST fixed indirectly with polyethylene terephthalate tape tied over a screw in a full outside-in created tunnel was superior to a ST/G graft fixed with an interference screw. Materials and methods: In a controlled laboratory study, six cadaveric matched pairs of each construct were subjected to cyclic loading to mimic physiologic loading during rehabilitation. This included preconditioning cycling, cyclic loading to 220 N for 500 cycles, then cyclic loading to 500 N for 500 cycles. Results: High standard deviations across the measured parameters occurred with no significant difference between measured parameters of elongation for the different constructs. Elongation of the Quad-ST construct was greater at 10 and 100 cycles, but not statistically different. Four of the six Quad-ST constructs failed below 100 cycles, compared with two failures below 100 cycles in the ST/G construct. There was a strong correlation between cycles to failure and bone mineral density for the Quad ST-tape constructs. Conclusions: Tibial fixation of Quad ST with a tied tape–screw construct in a full-length tunnel was not biomechanically superior to ST/G graft fixed with an interference screw, exhibited greater nonsignificant construct elongation with earlier failure, and was more reliant on bone mineral density.
View less >
View more >Background: In anterior cruciate ligament reconstruction, quadrupled semitendinosus (Quad ST) grafts have potential advantages over doubled semitendinosus–gracilis (ST/G) including larger diameter and gracilis preservation, however the ideal tibial fixation method of the resultant shorter Quad ST graft remains elusive if a fixed-loop suspensory fixation device is used on the femur. We investigated whether the tibial fixation biomechanical properties of a Quad ST fixed indirectly with polyethylene terephthalate tape tied over a screw in a full outside-in created tunnel was superior to a ST/G graft fixed with an interference screw. Materials and methods: In a controlled laboratory study, six cadaveric matched pairs of each construct were subjected to cyclic loading to mimic physiologic loading during rehabilitation. This included preconditioning cycling, cyclic loading to 220 N for 500 cycles, then cyclic loading to 500 N for 500 cycles. Results: High standard deviations across the measured parameters occurred with no significant difference between measured parameters of elongation for the different constructs. Elongation of the Quad-ST construct was greater at 10 and 100 cycles, but not statistically different. Four of the six Quad-ST constructs failed below 100 cycles, compared with two failures below 100 cycles in the ST/G construct. There was a strong correlation between cycles to failure and bone mineral density for the Quad ST-tape constructs. Conclusions: Tibial fixation of Quad ST with a tied tape–screw construct in a full-length tunnel was not biomechanically superior to ST/G graft fixed with an interference screw, exhibited greater nonsignificant construct elongation with earlier failure, and was more reliant on bone mineral density.
View less >
Journal Title
JOURNAL OF ORTHOPAEDICS AND TRAUMATOLOGY
Volume
19
Issue
1
Copyright Statement
© The Author(s) 2018. This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
Subject
Biomedical and clinical sciences