Show full item recordAbstract
Power quality and energy efficiency are of great importance in motor control. The motor of any medical device needs to have a smooth torque and minimal vibration in order to maximise its energy efficiency and patient comfort. Furthermore, in rotary blood pumps, excessive energy wasted due to vibration is converted into uncontrolled movement of the mechanical parts and thus could reduce the life of the motor-pump. Besides mechanical or hydraulic origin, one of the causes of vibration in any pump is torque ripple resulting from motor phase commutation. In this paper, using relevant equipment, two extreme scenarios were examined ...
View more >Power quality and energy efficiency are of great importance in motor control. The motor of any medical device needs to have a smooth torque and minimal vibration in order to maximise its energy efficiency and patient comfort. Furthermore, in rotary blood pumps, excessive energy wasted due to vibration is converted into uncontrolled movement of the mechanical parts and thus could reduce the life of the motor-pump. Besides mechanical or hydraulic origin, one of the causes of vibration in any pump is torque ripple resulting from motor phase commutation. In this paper, using relevant equipment, two extreme scenarios were examined for vibration and electrical efficiency comparison due to power quality in a blood pump: one trapezoidal control with a trapezoidal phase current output; the other a field oriented control (FOC) with a non-distorted sinusoidal phase current. The test motor-pump was the Arteriovenous Fistula Eligibility (AFE) System that is used prior to haemodialysis. The trapezoidal technique was implemented utilising the Allegro a4941 fan driver (Allegro Microsystem, 2012), and the FOC technique was implemented using the Texas Instrument digital signal processor (TMS320F28335). The aim was to reduce the energy wasted over vibration, and to achieve smooth operation of the AFE System. Vibration was measured with a one-axis accelerometer; results showed considerably lower vibration due to less current ripple associated with the FOC control as well as lower power consumption.
View less >
View more >Power quality and energy efficiency are of great importance in motor control. The motor of any medical device needs to have a smooth torque and minimal vibration in order to maximise its energy efficiency and patient comfort. Furthermore, in rotary blood pumps, excessive energy wasted due to vibration is converted into uncontrolled movement of the mechanical parts and thus could reduce the life of the motor-pump. Besides mechanical or hydraulic origin, one of the causes of vibration in any pump is torque ripple resulting from motor phase commutation. In this paper, using relevant equipment, two extreme scenarios were examined for vibration and electrical efficiency comparison due to power quality in a blood pump: one trapezoidal control with a trapezoidal phase current output; the other a field oriented control (FOC) with a non-distorted sinusoidal phase current. The test motor-pump was the Arteriovenous Fistula Eligibility (AFE) System that is used prior to haemodialysis. The trapezoidal technique was implemented utilising the Allegro a4941 fan driver (Allegro Microsystem, 2012), and the FOC technique was implemented using the Texas Instrument digital signal processor (TMS320F28335). The aim was to reduce the energy wasted over vibration, and to achieve smooth operation of the AFE System. Vibration was measured with a one-axis accelerometer; results showed considerably lower vibration due to less current ripple associated with the FOC control as well as lower power consumption.
View less >
Journal Title
Energies
Volume
13
Issue
15
Publisher URI
Copyright Statement
© 2020 The Authors. Licensee MDPI, Basel, Switzerland. This is an open access article distributed under the Creative Commons Attribution License which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited
Subject
Physical sciences
Engineering
Medical devices
Automation engineering
Built environment and design