The importance of low level QC for high sensitivity troponin assays
File version
Accepted Manuscript (AM)
Author(s)
Koerbin, Gus
Badrick, Tony
Oakman, Carmen
Potter, Julia M
Griffith University Author(s)
Primary Supervisor
Other Supervisors
Editor(s)
Date
Size
File type(s)
Location
Abstract
Background: With the advent of the new high-sensitivity troponin assays, it is becoming critical to measure troponin accurately to low concentrations. To ensure assay performance is acceptable, appropriate QC must be run. Methods: In addition to the routine use of commercial QC material, we prepared pools of human QC material with low troponin concentrations close to the limit of quantitation, and ran these regularly on our laboratory analysers. Results: Over 3 years we found no drift or shift in our hs-cTnI assay. We found that only the very low concentration human QC material gave warning of precision problems with the hs-cTnI assay. At the time of the documented poor assay precision, the higher concentration QC material indicated satisfactory performance. Conclusions: Choice of QC material with an appropriate concentration is important for any assay. For hs-cTn assays, it is of particular importance to use control material with a concentration near to the limit of quantitation.
Journal Title
Clinical Biochemistry
Conference Title
Book Title
Edition
Volume
58
Issue
Thesis Type
Degree Program
School
Publisher link
Patent number
Funder(s)
Grant identifier(s)
Rights Statement
Rights Statement
© 2018 Elsevier. Licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International Licence (http://creativecommons.org/licenses/by-nc-nd/4.0/) which permits unrestricted, non-commercial use, distribution and reproduction in any medium, providing that the work is properly cited.
Item Access Status
Note
Access the data
Related item(s)
Subject
Medical biochemistry and metabolomics
Clinical sciences
Science & Technology
Life Sciences & Biomedicine
Medical Laboratory Technology
Troponin
Limit of quantitation
Persistent link to this record
Citation
Hickmand, PE; Koerbin, G; Badrick, T; Oakman, C; Potter, JM, The importance of low level QC for high sensitivity troponin assays, Clinical Biochemistry, 2018, 58, pp. 60-63