Peripherally Inserted Central Catheter Thrombosis Following placement via electrocardiography vs. traditional methods: an analysis of 42,687 patients
File version
Accepted Manuscript (AM)
Author(s)
Horowitz, J
Rickard, CM
Ullman, AJ
Marsh, N
Schults, J
Ratz, D
Chopra, V
Griffith University Author(s)
Primary Supervisor
Other Supervisors
Editor(s)
Date
Size
File type(s)
Location
Abstract
Background: Peripherally inserted central catheter tip placement at the cavoatrial junction is associated with reduced catheter-related deep vein thrombosis . Electrocardiographic tip confirmation purportedly improves accuracy of tip placement, but whether this approach can reduce deep vein thrombosis is unknown. Methods: Prospectively collected data from patients that received peripherally inserted central catheters at 52 Michigan hospitals were analyzed. The method used to confirm tip confirmation at insertion and deep vein thrombosis outcomes were extracted from medical records. Multivariate models (accounting for the clustered nature of the data) were fitted to assess the association between peripherally inserted central catheter-related deep vein thrombosis and method of tip confirmation (Electrocardiographic vs. radiographic imaging). Results: A total of 42,687 peripherally inserted central catheters (21,098 radiology vs. 21,589 Electrocardiographic) were included. Patients receiving Electrocardiographic -confirmed peripherally inserted central catheters had fewer comorbidities compared to those that underwent placement via radiology. Overall, deep vein thrombosis occurred in 594 (1.3%) of all peripherally inserted central catheters. Larger catheter size (odds radio [OR]=1.32 [95% confidence interval [CI] 0.93-1.90] per unit increase in gauge), history of deep vein thrombosis, and cancer were associated with increased risk of deep vein thrombosis (OR=2.00 [95%CI 1.65-2.43] and OR=1.62 [95%CI 1.16-2.26], respectively) using logistic regression. Following adjustment, electrocardiographic guidance was associated with a significant reduction in peripherally inserted central catheter-related deep vein thrombosis compared with radiographic imaging (OR=0.74, 95% [CI=0.58-0.93, p=0.0098]). Conclusion: The use of electrocardiographic to confirm peripherally inserted central catheter tip placement at the cavoatrial junction was associated with significantly fewer deep vein thrombosis events than radiographic imaging. Use of this approach for peripherally inserted central catheter insertion may help improve patient safety, particularly in high risk patients.
Journal Title
American Journal of Medicine
Conference Title
Book Title
Edition
Volume
Issue
Thesis Type
Degree Program
School
Publisher link
Patent number
Funder(s)
Grant identifier(s)
Rights Statement
Rights Statement
© 2020 Elsevier. Licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International Licence, which permits unrestricted, non-commercial use, distribution and reproduction in any medium, providing that the work is properly cited.
Item Access Status
Note
This publication has been entered into Griffith Research Online as an Advanced Online Version.
Access the data
Related item(s)
Subject
Biomedical and clinical sciences
Clinical sciences
Catheter
EKG
PICC
Venous Thromboembolism
Persistent link to this record
Citation
Kleidon, TM; Horowitz, J; Rickard, CM; Ullman, AJ; Marsh, N; Schults, J; Ratz, D; Chopra, V, Peripherally Inserted Central Catheter Thrombosis Following placement via electrocardiography vs. traditional methods: an analysis of 42,687 patients., American Journal of Medicine, 2020