HeartWare HVAD Flow Estimator Accuracy for Left and Right Ventricular Support

No Thumbnail Available
File version
Author(s)
Stephens, Andrew F
Mapley, Martin
Wu, Eric L
Fraser, John F
Steinseifer, Ulrich
Bartnikowski, Nicole
Gregory, Shaun D
Griffith University Author(s)
Primary Supervisor
Other Supervisors
Editor(s)
Date
2021
Size
File type(s)
Location
License
Abstract

This study investigated the accuracy of the HeartWare HVAD flow estimator for left ventricular assist device (LVAD) support and biventricular assist device (BiVAD) support for modes of reduced speed (BiVAD-RS) and banded outflow (BiVAD-B). The HVAD flow estimator was evaluated in a mock circulatory loop under changes in systemic and pulmonary vascular resistance, heart rate, central venous pressure, and simulated hematocrit (correlated to viscosity). A difference was found between mean estimated and mean measured flow for LVAD (0.1 ± 0.3 L/min), BiVAD-RS (-0.1 ± 0.2 L/min), and BiVAD-B (0 ± 0.2 L/min). Analysis of the flow waveform pulsatility showed good correlation for LVAD (r2= 0.98) with a modest spread in error (0.7 ± 0.1 L/min), while BiVAD-RS and BiVAD-B showed similar spread in error (0.7 ± 0.3 and 0.7 ± 0.2 L/min, respectively), with much lower correlation (r2= 0.85 and r2= 0.60, respectively). This study demonstrated that the mean flow error of the HVAD flow estimator is similar when the device is used in LVAD, BiVAD-RS, or BiVAD-B configuration. However, the instantaneous flow waveform should be interpreted with caution, particularly in the cases of BiVAD support.

Journal Title

ASAIO journal

Conference Title
Book Title
Edition
Volume

67

Issue

4

Thesis Type
Degree Program
School
Publisher link
Patent number
Funder(s)
Grant identifier(s)
Rights Statement
Rights Statement
Item Access Status
Note
Access the data
Related item(s)
Subject

Biomedical engineering

Science & Technology

Life Sciences & Biomedicine

Transplantation

Persistent link to this record
Citation

Stephens, AF; Mapley, M; Wu, EL; Fraser, JF; Steinseifer, U; Bartnikowski, N; Gregory, SD, HeartWare HVAD Flow Estimator Accuracy for Left and Right Ventricular Support, ASAIO journal, 2021, 67 (4), pp. 416-422