Assay Stability, the missing component of the Error Budget

No Thumbnail Available
File version
Author(s)
Mackay, Mark
Hegedus, Gabe
Badrick, Tony
Griffith University Author(s)
Primary Supervisor
Other Supervisors
Editor(s)
Date
2017
Size
File type(s)
Location
License
Abstract

Gaining a better understanding of Quality Control (QC) processes is a key requirement to improving performance and reducing patient risk. Detecting analytical error is dependent on a QC strategy that reliably detects a critical shift in a result away from the true value. Recently the concept of Six Sigma has been used by diagnostic laboratories to assess the performance of assays and to assist in the selection of QC rules. The sigma metric is one measure of an assay's ability to perform within specification. However an additional dimension to managing an assay is its stability in bias over time. The concept of long term stability is the same as measured QC drift (SEdrift) which is the effect of numerous calibrations, changes in reagent lots and other conditions i.e. a long term effect. This implies that the standard error budget is wrong because it is modelled on short term QC and misses this SEdrift stability component. We show that SEdrift provides a measure of Assay Stability that should be included in Quality Planning and that by including an allowance for this drift, determining target imprecision appropriate for matched QC algorithms that provide high error detection is as simple as dividing the Allowable Performance Specification by 4, 5 or 6.

Journal Title
Clinical Biochemistry
Conference Title
Book Title
Edition
Volume
Issue
Thesis Type
Degree Program
School
Publisher link
Patent number
Funder(s)
Grant identifier(s)
Rights Statement
Rights Statement
Item Access Status
Note
Access the data
Related item(s)
Subject
Medical biochemistry and metabolomics
Clinical sciences
Science & Technology
Life Sciences & Biomedicine
Medical Laboratory Technology
Error detection
QC rules
Persistent link to this record
Citation
Mackay, M; Hegedus, G; Badrick, T, Assay Stability, the missing component of the Error Budget, Clinical Biochemistry, 2017, 50 (18), pp. 1136-1144