Improving The Accuracy Of A Turbine Spirometer At Low Flow Rates

No Thumbnail Available
File version
Author(s)
Cross, Troy J
Isautier, Jennifer MJ
Ziegler, Briana L
Cierzan, Bradley S
Wentz, Robert
Carlson, Alex
Johnson, Bruce D
Griffith University Author(s)
Primary Supervisor
Other Supervisors
Editor(s)
Date
2018
Size
File type(s)
Location

Minneapolis, MN

License
Abstract

The turbine spirometer is a popular laboratory device used to measure respiratory volumes and flows during exercise, and while performing gross, voluntary respiratory manoeuvres (i.e., inspiratory capacity efforts). Indeed, these devices have been incorporated into many commercially-available pulmonary function and metabolic systems. Yet, while the turbine spirometer may provide accurate/reliable measurements of respiratory volumes at modest-to-high flows, these devices perform poorly at low flow rates.

PURPOSE: To improve the accuracy of a turbine spirometer over an extended range of low flows using the “weighted averaging technique” described by Yeh et al. (J Appl Physiol, 53(1): p280, 1982).

METHODS: A commercially-available turbine spirometer was interfaced with a custom-designed microcontroller unit (MCU). The MCU recorded discrete rotations of the turbine rotor, and the corresponding rotational frequency (frot). Repeated 5-fold cross-validation was used to determine the optimal number of bins in frot and iterations used in the Weighted Averaging algorithm. This method yielded a discrete array of calibration constants (K) across a relevant range of frot. (<1-1800 Hz). The accuracy of this “nonlinear” calibration curve was compared to that obtained by assuming a constant K across all frequencies (i.e., flows). Over 200 calibrations strokes were recorded using a 3 L syringe.

RESULTS: By assuming a constant K (15.6 mL[BULLET OPERATOR]pulse-1), the turbine spirometer exhibited an average volume error of +94 mL (+3.1%) over a 95% confidence interval (CI95%) of -856 to +375 mL (-28.5 to 12.5%). Conversely, applying the nonlinear K curve resulted in an average volume error of <0.001 mL (<0.001%) and a CI95% ranging from -60 to +60 mL (-2.0 to 2.0%). Importantly, the nonlinear K curve provided accurate (within ±3%) volume measurements down to 0.33 Hz (~7 mL[BULLET OPERATOR]s-1).

CONCLUSIONS: The “weighted averaging technique” improved the accuracy/reliability the turbine spirometer to within ±3% across an interval of flows ranging between ~0.01 to 20 L[BULLET OPERATOR]s-1.

Journal Title
Conference Title

Medicine & Science in Sports & Exercise

Book Title
Edition
Volume

50

Issue

5S

Thesis Type
Degree Program
School
Publisher link
Patent number
Funder(s)
Grant identifier(s)
Rights Statement
Rights Statement
Item Access Status
Note
Access the data
Related item(s)
Subject

Sports science and exercise

Medical physiology

Science & Technology

Life Sciences & Biomedicine

Sport Sciences

Persistent link to this record
Citation

Cross, TJ; Isautier, JMJ; Ziegler, BL; Cierzan, BS; Wentz, R; Carlson, A; Johnson, BD, Improving The Accuracy Of A Turbine Spirometer At Low Flow Rates, Medicine & Science in Sports & Exercise, 2018, 50 (5S), pp. 286-286