Midpregnancy testing for soluble fms-like tyrosine kinase 1 (sFlt-1) and placental growth factor (PlGF): An inter-assay comparison of three automated immunoassay platforms

No Thumbnail Available
File version
Author(s)
Black, Carin
Al-Amin, Ahmed
Rolnik, Daniel Lorber
Kane, Stefan C
Stolarek, Caroline
White, Adrienne
Costa, Fabricio da Silva
Brennecke, Shaun
Griffith University Author(s)
Primary Supervisor
Other Supervisors
Editor(s)
Date
2019
Size
File type(s)
Location
License
Abstract

We performed an inter-assay comparison among three immunoassay platforms for midpregnancy testing of sFlt-1, PlGF and the sFlt-1/PlGF ratio, which are established markers for pre-eclampsia. Maternal blood was collected 19–22 weeks’ gestation. Raw data values were converted to multiples of the median (MoM). PlGF and sFlt-1 values among platforms were highly correlated (p < 0.001). There was significant variation in raw data values for PlGF and sFlt-1 among platforms, eliminated following conversion to MoM. When directly comparing raw data values among platforms, platform-specific reference ranges values should be used. MoM values were equivalent among platforms, allowing direct inter-assay result comparison.

Journal Title

Placenta

Conference Title
Book Title
Edition
Volume

86

Issue
Thesis Type
Degree Program
School
Publisher link
Patent number
Funder(s)
Grant identifier(s)
Rights Statement
Rights Statement
Item Access Status
Note
Access the data
Related item(s)
Subject

Biochemistry and cell biology

Clinical sciences

Science & Technology

Life Sciences & Biomedicine

Developmental Biology

Obstetrics & Gynecology

Reproductive Biology

Persistent link to this record
Citation

Black, C; Al-Amin, A; Rolnik, DL; Kane, SC; Stolarek, C; White, A; Costa, FDS; Brennecke, S, Midpregnancy testing for soluble fms-like tyrosine kinase 1 (sFlt-1) and placental growth factor (PlGF): An inter-assay comparison of three automated immunoassay platforms, Placenta, 2019, 86, pp. 11-14

Collections