Review of current incidents and risk calculations used in the Royal College of Australasian Pathologists Key Incident Management and Monitoring Systems - a system that could be used by all Australasian medical laboratories, and easily adapted to worldwide use

Loading...
Thumbnail Image
File version

Version of Record (VoR)

Author(s)
Gay, Stephanie
Pope, Belinda
Badrick, Tony
Whiley, Michael
Griffith University Author(s)
Primary Supervisor
Other Supervisors
Editor(s)
Date
2022
Size
File type(s)
Location
Abstract

Introduction: The Royal College of Pathologists of Australasia Quality Assurance Programs (RCPAQAP) Key Incident Monitoring and Management Systems (KIMMS) program has found that some existing Quality Indicators are too broad or not well defined. The risk matrix in use does not allow changes in incident Detection or Probability. In 2020, a review was performed: what issues should KIMMS include as Key Incidents and how could risk measurement be improved? Materials and methods: Twenty-seven networked and stand-alone laboratories enrolled in KIMMS during 2020 were surveyed on 45 current and new indicators of risk in the total testing process. They were asked which indicators they considered were significant in causing patient harm. Existing risk matrices in use by members of the KIMMS Advisory Committee laboratories were reviewed regarding their size or structure (3x3 or 5x5) and the descriptions of consequences and probability. Results: Thirteen participants indicated 21 indicators should be monitored, and the KIMMS Advisory committee added a further 13 (11 from the remaining 24 and 2 new). Of the five risk matrices reviewed, all consistently used a 5x5 matrix to estimate Consequences vs Probability of harm. The KIMMS advisory committee added a third parameter to the calculation of Risk, Detectability. Conclusion: All 34 pre-and post-indicators should be monitored, covering all aspects of the total testing cycle other than analytical. The risk measurement can be improved by introducing a 5x5 risk matrix to evaluate harm (consequences x probability) and then evaluating risk by adding detectability; risk equals harm x detectability.

Journal Title

Biochemia Medica

Conference Title
Book Title
Edition
Volume

32

Issue

1

Thesis Type
Degree Program
School
Publisher link
Patent number
Funder(s)
Grant identifier(s)
Rights Statement
Rights Statement

© 2021, Croatian Society of Medical Biochemistry. This work is licensed under a Creative Commons Attribution 4.0 International License.

Item Access Status
Note
Access the data
Related item(s)
Subject

Patient safety

Health services and systems

Quality management

Clinical microbiology

Health systems

Science & Technology

Life Sciences & Biomedicine

Medical Laboratory Technology

extra-analytical phase

laboratory organisation and management

Persistent link to this record
Citation

Gay, S; Pope, B; Badrick, T; Whiley, M, Review of current incidents and risk calculations used in the Royal College of Australasian Pathologists Key Incident Management and Monitoring Systems - a system that could be used by all Australasian medical laboratories, and easily adapted to worldwide use, Biochemia Medica, 2022, 32 (1), pp. 010702

Collections