Mitral Valve Regurgitation with a Rotary Left Ventricular Assist Device: The Haemodynamic Effect of Inlet Cannulation Site and Speed Modulation
File version
Author(s)
Stevens, Michael C
Wu, Eric L
Pauls, Jo P
Kleinheyer, Matthias
Fraser, John F
Griffith University Author(s)
Primary Supervisor
Other Supervisors
Editor(s)
Date
Size
File type(s)
Location
License
Abstract
Mitral valve regurgitation (MVR) is common in patients receiving left ventricular assist device (LVAD) support, however the haemodynamic effect of MVR is not entirely clear. This study evaluated the haemodynamic effect of MVR with LVAD support and the influence of inflow cannulation site and LVAD speed modulation. Left atrial (LAC) and ventricular (LVC) cannulation was evaluated in a mock circulation loop with no, mild, moderate and severe MVR with constant speed and speed modulation (±600 RPM) modes. The use of an LVAD relieved pulmonary congestion during severe MVR, by reducing left atrial pressure from 20.5 to 10.8 (LAC) and 11.5 (LVC) mmHg. However, LAC resulted in decreased left ventricular stroke work (−0.08 J), ejection fraction (−7.9%) and higher MVR volume (+12.7 mL) and pump speed (+100 RPM) compared to LVC. This suggests that LVC, in addition to reducing MVR severity, also improves ventricular washout over LAC. LVAD speed modulation in synchrony with ventricular systole reduced MVR volume and increased ejection fraction with LAC and LVC, thus demonstrating the potential benefits of this mode, despite a reduction in cardiac output.
Journal Title
Annals of Biomedical Engineering
Conference Title
Book Title
Edition
Volume
44
Issue
9
Thesis Type
Degree Program
School
Publisher link
Patent number
Funder(s)
Grant identifier(s)
Rights Statement
Rights Statement
Item Access Status
Note
Access the data
Related item(s)
Subject
Engineering
Medical devices
Biomedical and clinical sciences