Abstract
A radiation sterilization dose (RSD) of 25 kGy is deleterious to bone allografts. This study aimed to establish a lower RSD for bone allografts using method 1 of International Standard Organisation 11137.2:2006. This provides a database to select an RSD corresponding to an allograft's bioburden, given that the bioburden's gamma resistance is equal to or less than the standard. This can be verified by irradiating 100 allografts at a dose selected to provide a sterility assurance level of 10-2. The bioburden of our allografts was 0, which prescribed a verification dose of 1.3 kGy. After irradiating 100 llografts, sterility tests returned no positive cultures. We therefore validated an RSD of 11 kGy for allografts with that bioburden. According to the standard, this RSD provides a sterility assurance level of 10-6 for bone allografts.
Journal Title
Journal of Arthroplasty
Conference Title
Book Title
Edition
Volume
26
Issue
2
Thesis Type
Degree Program
School
Publisher link
Patent number
Funder(s)
Grant identifier(s)
Rights Statement
Rights Statement
© 2010 Elsevier Inc. This is the author-manuscript version of this paper. Reproduced in accordance with the copyright policy of the publisher. Please refer to the journal's website for access to the definitive, published version.
Item Access Status
Note
Access the data
Related item(s)
Subject
Biomedical engineering
Clinical sciences
Orthopaedics