Ex vivo assessment of erythrocyte tolerance to the HeartWare ventricular assist device operated in three discrete configurations

Loading...
Thumbnail Image
File version

Accepted Manuscript (AM)

Author(s)
Kuck, Lennart
Simmonds, Michael J
Chan, Chris Hoi Houng
Pauls, Jo P
Tansley, Geoff D
Feldmann, Friederike
McNamee, Antony P
Primary Supervisor
Other Supervisors
Editor(s)
Date
2020
Size
File type(s)
Location
License
Abstract

Despite technological advances in ventricular assist devices (VADs) to treat end-stage heart failure, hemocompatibility remains a constant concern, with supraphysiological shear stresses an unavoidable reality with clinical use. Given that impeller rotational speed is related to the instantaneous shear within the pump housing, it is plausible that modulation of pump speed may regulate peak mechanical shear stresses and thus ameliorate blood damage. The present study investigated the hemocompatibility of the HeartWare HVAD in three configurations typical of clinical applications: standard systemic support left VAD (LVAD), pediatric support LVAD, and pulmonary support right VAD (RVAD) conditions. Two ex vivo mock circulation blood loops were constructed using explanted HVADs, in which pump speed and external loop resistance were manipulated to reflect the flow rates and differential pressures reported in configurations for standard adult LVAD (at 3150 rev⸱min-1 ), pediatric LVAD (at 2400 rev⸱min-1 ), and adult RVAD (at 1900 rev⸱min-1 ). Using bovine blood, the mock circulation blood loops were tested at 37°C over a period of 6 hours (consistent with ASTM F1841-97) and compared with a static control. Hemocompatibility assessments were conducted for each test condition, examining hematology, hemolysis (absolute and normalized index), osmotic fragility, and blood viscosity. Regardless of configuration, continuous exposure of blood to the VAD over the 6-hour period significantly altered hematological and rheological blood parameters, and induced increased hemolysis when compared with a static control sample. Comparison of the three operational VAD configurations identified that the adult LVAD condition - associated with the highest pump speed, flow rate, and differential pressure across the pump - resulted in increased normalized hemolysis index (NIH; 0.07) when compared with the lower pump speed 'off-label' counterparts (NIH of 0.04 in pediatric LVAD and 0.01 in adult RVAD configurations). After normalizing blood residence times between configurations, pump speed was identified as the primary determinant of accumulated blood damage; plausibly, blood damage could be limited by restricting pump speed to the minimum required to support matched cardiac output, but not beyond.

Journal Title

Artificial Organs

Conference Title
Book Title
Edition
Volume
Issue
Thesis Type
Degree Program
School
Publisher link
Patent number
Funder(s)
Grant identifier(s)
Rights Statement
Rights Statement

© 2020 International Center for Artificial Organs and Transplantation and Wiley Periodicals, Inc. This is the peer reviewed version of the following article: Ex vivo assessment of erythrocyte tolerance to the HeartWare ventricular assist device operated in three discrete configurations, Artificial Organs, 2020, which has been published in final form at https://doi.org/10.1111/aor.13877. This article may be used for non-commercial purposes in accordance with Wiley Terms and Conditions for Self-Archiving (http://olabout.wiley.com/WileyCDA/Section/id-828039.html)

Item Access Status
Note

This publication has been entered as an advanced online version in Griffith Research Online.

Access the data
Related item(s)
Subject

Biomedical engineering

Clinical sciences

blood damage

blood viscosity

pediatric support

pulmonary support

systemic support

Persistent link to this record
Citation

Kuck, L; Simmonds, MJ; Chan, CHH; Pauls, JP; Tansley, GD; Feldmann, F; McNamee, AP, Ex vivo assessment of erythrocyte tolerance to the HeartWare ventricular assist device operated in three discrete configurations., Artificial Organs, 2020

Collections