Understanding Patient-Based Real-Time Quality Control Using Simulation Modeling.

No Thumbnail Available
File version
Author(s)
Bietenbeck, Andreas
Cervinski, Mark A
Katayev, Alex
Loh, Tze Ping
van Rossum, Huub H
Badrick, Tony
Griffith University Author(s)
Primary Supervisor
Other Supervisors
Editor(s)
Date
2020
Size
File type(s)
Location
License
Abstract

Background: Patient-based real-time quality control (PBRTQC) avoids limitations of traditional quality control methods based on the measurement of stabilized control samples. However, PBRTQC needs to be adapted to the individual laboratories with parameters such as algorithm, truncation, block size, and control limit. Methods: In a computer simulation, biases were added to real patient results of 10 analytes with diverse properties. Different PBRTQC methods were assessed on their ability to detect these biases early. Results: The simulation based on 460 000 historical patient measurements for each analyte revealed several recommendations for PBRTQC. Control limit calculation with "percentiles of daily extremes" led to effective limits and allowed specification of the percentage of days with false alarms. However, changes in measurement distribution easily increased false alarms. Box-Cox but not logarithmic transformation improved error detection. Winsorization of outlying values often led to a better performance than simple outlier removal. For medians and Harrell-Davis 50 percentile estimators (HD50s), no truncation was necessary. Block size influenced medians substantially and HD50s to a lesser extent. Conversely, a change of truncation limits affected means and exponentially moving averages more than a change of block sizes. A large spread of patient measurements impeded error detection. PBRTQC methods were not always able to detect an allowable bias within the simulated 1000 erroneous measurements. A web application was developed to estimate PBRTQC performance. Conclusions: Computer simulations can optimize PBRTQC but some parameters are generally superior and can be taken as default.

Journal Title

Clinical Chemistry

Conference Title
Book Title
Edition
Volume

66

Issue

8

Thesis Type
Degree Program
School
Publisher link
Patent number
Funder(s)
Grant identifier(s)
Rights Statement
Rights Statement
Item Access Status
Note
Access the data
Related item(s)
Subject

Medical biotechnology

Medical biochemistry and metabolomics

Clinical sciences

average of normals

exponentially weighted moving average

moving average

optimization

quality control

Persistent link to this record
Citation

Bietenbeck, A; Cervinski, MA; Katayev, A; Loh, TP; van Rossum, HH; Badrick, T, Understanding Patient-Based Real-Time Quality Control Using Simulation Modeling., Clinical Chemistry, 2020, 66 (8), Pages 1072–1083

Collections