Stress Evaluation of Dental Implant Wall Thickness using Numerical Techniques
No Thumbnail Available
File version
Author(s)
Van Staden, Rudi
Guan, Hong
Loo, Yew-Chaye
Johnson, Newell
Meredith, Neil
Guan, Hong
Loo, Yew-Chaye
Johnson, Newell
Meredith, Neil
Griffith University Author(s)
Primary Supervisor
Other Supervisors
Editor(s)
Tomas Albrektsson
Date
2008
Size
File type(s)
Location
License
Abstract
Using the Finite Element Method, four implant diameters were evaluated for their effect on the stress distribution at the implant wall. A two-dimensional model of the implant and mandibular bone used triangular and quadrilateral plane strain elements to compute the von Mises stress in the bone with varied masticatory forces and abutment screw preloads. The masticatory force is found to be more influential than abutment screw preload, and implant wall thickness significantly influences the stress magnitude within the implant.
Journal Title
Applied Osseointegration Research
Conference Title
Book Title
Edition
Volume
6