Global tidal variations, regional distribution of ventilation, and the regional onset of filling determined by electrical impedance tomography: reproducibility

Loading...
Thumbnail Image
File version
Author(s)
Caruana, LR
Barnett, AG
Tronstad, O
Paratz, JD
Chang, AT
Fraser, JF
Griffith University Author(s)
Primary Supervisor
Other Supervisors
Editor(s)
Date
2017
Size
File type(s)
Location
License
Abstract

The reproducibility of the regional distribution of ventilation and the timing of onset of regional filling as measured by electrical impedance tomography lacks evidence. This study investigated whether electrical impedance tomography measurements in healthy males were reproducible when electrodes were replaced between measurements. Part 1: Recordings of five volunteers lying supine were made using electrical impedance tomography and a pneumotachometer. Measurements were repeated at least three hours later. Skin marking ensured accurate replacement of electrodes. No stabilisation period was allowed. Part 2: Electrical impedance tomography recordings of ten volunteers; a 15 minute stabilisation period, extra skin markings, and time-averaging were incorporated to improve the reproducibility. Reproducibility was determined using the Bland-Altman method. To judge the transferability of the limits of agreement, a Pearson correlation was used for electrical impedance tomography tidal variation and tidal volume. Tidal variation was judged to be reproducible due to the significant correlation between tidal variation and tidal volume (r2 = 0.93). The ventilation distribution was not reproducible. A stabilisation period, extra skin markings and time-averaging did not improve the outcome. The timing of regional onset of filling was reproducible and could prove clinically valuable. The reproducibility of the tidal variation indicates that non-reproducibility of the ventilation distribution was probably a biological difference and not measurement error. Other causes of variability such as electrode placement variability or lack of stabilisation when accounted for did not improve the reproducibility of the ventilation distribution.

Journal Title

Anaesthesia and Intensive Care

Conference Title
Book Title
Edition
Volume

45

Issue

2

Thesis Type
Degree Program
School
DOI
Patent number
Funder(s)
Grant identifier(s)
Rights Statement
Rights Statement

© 2017 Australian Society of Anaesthetists. The attached file is reproduced here in accordance with the copyright policy of the publisher. Please refer to the journal's website for access to the definitive, published version.

Item Access Status
Note
Access the data
Related item(s)
Subject

Biomedical and clinical sciences

Persistent link to this record
Citation
Collections