Introduction: The capability of glucometer internal quality control (QC) in detecting varying magnitude of systematic error (bias), and the potential use of moving sum of positive results (MovSum) and moving average (MA) techniques as potential alternatives were evaluated. Materials and methods: The probability of error detection using routine QC and manufacturer’s control limits were investigated using historical data. Moving sum of positive results and MA algorithms were developed and optimized before being evaluated through numerical simulation for false positive rate and probability of error detection. Results: When the manufacturer’s default control limits (that are multiple times higher than the running standard deviation (SD) of the glucome-ter) was used, they had 0-75% probability of detecting small errors up to 0.8 mmol/L. However, the error detection capability improved to 20-100% when the running SD of the glucometer was used. At a binarization threshold of 6.2 mmol/L and block sizes of 200 to 400, MovSum has a 100% probability of detecting a bias that is greater than 0.5 mmol/L. Compared to MovSum, the MA technique had lower probability of bias detection, especially for smaller bias magnitudes; MA also had higher false positive rates. Conclusions: The MovSum technique is suited for detecting small, but clinically significant biases. Point of care QC should follow conventional practice by setting the control limits according to the running mean and SD to allow proper error detection. The glucometer manufacturers have an active role to play in liberalizing QC settings and also enhancing the middleware to facility patient-based QC practices.
Journal Title
Biochemia Medica
Conference Title
Book Title
Edition
Volume
30
Issue
2
Thesis Type
Degree Program
School
Publisher link
Patent number
Funder(s)
Grant identifier(s)
Rights Statement
Rights Statement
© Croatian Society of Medical Biochemistry and Laboratory Medicine. This work is licensed under a Creative Commons Attribution 4.0 International License which permits users to read, download, copy, distribute, print, search, or link to the full texts of these articles in any medium or format and to remix, transform and build upon the material, provided the original work is properly cited and any changes properly indicated.
Item Access Status
Note
Access the data
Related item(s)
Subject
Medical biochemistry and metabolomics
Science & Technology
Life Sciences & Biomedicine
Medical Laboratory Technology
laboratory management
quality control
Persistent link to this record
Citation
Lim, CY; Badrick, T; Loh, TP, Patient-based quality control for glucometers: using the moving sum of positive patient results and moving average, BIOCHEMIA MEDICA, 2020, 30 (2)